Rf communication and signals experiments – GW Instek GRF-1300A User Manual
Page 53

RF Communication and Signals Experiments
marker function on the spectrum analyzer to measure the
value.
Step6
Marker
1
0
Record the value, then calculate the phase noise according to
the formula, and record the spectrum and measurement results
in Table 4-3.
6. Adjust the PLL output frequency to 900MHz, and again
measure the power and phase noise corresponding to the
frequency.
Step7
UP
Step8
F1
0
Step9
Peak
Search
Record the carrier power. Set the deviation carrier frequency
f
m
to a deviation ( of 100kHz. Use the Delta Marker function on
the spectrum analyzer to measure the value.
Step10
Marker
1
0
Record the value, then calculate the phase noise according
to the formula, and record the spectrum and measurement
results in Table 4-3.
7. Adjust the PLL output frequency to 910MHz, and again
measure the power and phase noise corresponding to the
frequency.
Step11
UP
51