Save/recall trace data – GW Instek GSP-730 User Manual User Manual
Page 74

GSP-730 User Manual
72
Save/Recall Trace Data
Description
The trace data can be saved/recalled for any of
the A, B or C traces to/from one of 5 pre-set
internal memory locations. The trace data
cannot be recalled or saved to USB.
When saving or recalling trace data from a split
spectrum, only the active spectrum is
saved/recalled.
Save
1. To save the current trace data, press
Memory
>
Save Trace Data..[F3]
2. Press Source Trace[F1] and select the source:
Source:
A, B, C
3. Press Destination[F2] and select the memory
location to save to:
Destination:
1~5
4. Press Start[F5] to save the selected tracedata.
Recall
5. To recall trace data, press
Memory
> Recall Trace
Data..[F3]
6. Press Source Trace[F1] and select the memory
location to recall from:
Source:
1~5
7. Press Destination[F2] and select the destination
trace
Destination:
A, B, C