Remote control function check – GW Instek GSP-730 User Manual User Manual
Page 82

GSP-730 User Manual
80
Panel operation
1. USB Connection:
Connect a USB cable from the PC
to the rear panel USB B port.
RS232 Connection:
Connect an RS232C cable from the
PC to the rear panel RS232 port.
2. Press
System
>Serial Port…[F1]> Serial[F1] to
enter the remote configuration.
3. Set the following RS-232 settings using the
arrow keys:
Baud Rate[F1]:
9600, 19200, 38400, 57600,
115200.
Parity[F2]:
None, Even, Odd, Space,
Mark, Multidrop.
Stop Bit[F3]:
1, 1.5, 2.
Data[F4]:
5, 6, 7, 8
Remote Control Function Check
Functionality
check
Invoke a terminal application such as MTTTY
(Multi-Threaded TTY).
To check the COM port No., see the Device
Manager in the PC.
For WinXP go to;
Control panel → System → Hardware tab.
Run this query command via the terminal after
the instrument has been configured for remote
control (page 79).
*idn?