Example
:SENSe:PCURrent:TIME:LOW?
Returns the integration time for low pulse
measurement.
Command
:SENSe[1]:PCURrent:TIME:AVERage <NRf>
Function
Sets the integration time for the average pulse
measurement.
Description
NRf
33-833333, step resolution of 33.3
Example
:SENSe:PCURrent:TIME:AVERage 0.000233
Sets the integration time for average pulse
measurement to 233 microseconds.
Command
:SENSe[1]:PCURrent:TIME:AVERage?
Function
Returns the integration time for the average
measurement.
Example
:SENSe:PCURrent:TIME:AVERage?
Returns the integration time for the average
measurement.
Command
:SENSe[1]:PCURrent:SYNChronize[:STATe] <b>
Function
Sets the triggering option for pulse current
measurement.
Description
<b>
0 /OFF: Digital trigger mode.
1/ON: Pulse level trigger mode.
Example
:SENSe:PCURrent:SYNChronize ON
The trigger mode is set to the pulse level trigger.