Readback commands – GW Instek PPH-1503 User Manual
Page 95

REMOTE CONTROL
95
Readback Commands
Command
:SENSe[1]:FUNCtion <name>
Function
Selects the type of measurement function: voltage,
current, pulse, long integration and DVM
measurement.
Description
name
“VOLTage”: Voltage measurement.
“CURRent”: Current measurement.
“PCURrent”: Pulse current
measurement.
“LINTegration”: Long integration
measurement.
“DVMeter”: DVM input measurement.
Example
:SENSe:FUNCtion “VOLTage”
Selects “Voltage” as the measurement type.
Command
:SENSe[1]:FUNCtion?
Function
Queries the type of measurement function.
Response time Maximum: 16ms
Example
:SENSe:FUNCtion?
Returns the type of measurement function.
Command
:SENSe[1]:NPLCycles <n>
Function
Sets the number of PLCs for the integration rate
for voltage, current and DVM measurements.
Description
<n>
0.01-10.00
Example
:SENSe:NPLCycles 0.10
Sets the number of PLCs to 0.1.