GW Instek GSP-827 User Manual User Manual

Page 18

Advertising
background image

2.7GHz Spectrum Analyzer Operation Manual

17

6.2.7 Delta-Marker operation

Function key

Description

Marker

Select the Marker relative functions.

F1

<Marker>

Select the active marker by keying in marker number. 0

means marker 10.

1

Key in 1 to take marker 1 as example.

F2 <Marker

ON/OFF>

Set Marker 1 ON.

N, … , MHz

Set frequency of marker 1.

F1

<Marker>

Select the second marker, marker 2 here for example.

2

Key in 2 to take marker 2 as example.

F2 <Marker

ON/OFF>

Set Marker 2 ON.

N, … , MHz

Set frequency of marker 2.

F3 <Normal,

Δ

Mkr>

Set Marker2 as Δ marker.

1

Select the Reference Marker due to Mkr2. Key in 1 to take

marker 1 as reference marker.

Note: The table will show the level and frequency in delta reading. In this example, the

marker table will show as follows.

Marker Level Freq(MHz)

1 l1 f1

3

Δ

1

Δ

l13

Δ

f13

Which means marker 3 is the ΔMkr due to Mkr1.

6.2.8 Peak Hold, Average and Freeze the measurement

Function key

Description

Trace

Select the trace relative functions.

F3 <PkHold>

Switch the peak hold function ON/OFF.

F4

<AVG >

Switch the Average function ON/OFF. When it is ON, key

in the average times.

F5

<Freeze>

Switch the freeze function ON/OFF.

6.2.9 ACPR Measurement

Function key

Description

Pwr Measure

Select the power measurement relative functions.

F4

<Setup>

Setup the Power Measurement parameters.

F1

<CH BW>

Define the channel bandwidth.

F2

<CH SPC>

Define the channel space.

F4<ADJ CH

OFFSET>

Define the adjacent channel parameters.

Advertising