Trigger, Free run, Trigger video level – GW Instek GSP-827 User Manual User Manual
Page 76: 7ghz spectrum analyzer operation manual ⎯ ⎯ 75

2.7GHz Spectrum Analyzer Operation Manual
⎯
⎯
75
Exter
Exter
Free Run
Free Run will stop trigger function and stay in normal sweep mode.
Video Level
UP/DOWN
and Scroll: To change the video level by 0.1dBm for every movement.
Editing
: To enter the level on display as the trigger threshold.
Example: Video level is set at –40dBm. The trace freezes when the trigger condition is not met.
When the input signal level is over –40dBm, the sweep starts and captures the signal.
1
Trigger
Start:99.5MHz
RBW:30 kHz
Stop:100.5MHz
VBW:100 kHz
Center:100.0MHz
Span:1.0MHz
Sweep:100.0ms
Video Level:-40 dBm
Free
Run
Ref
Lvl: -30.0 dBm
-30.0
-40.0
-50.0
-60.0
-70.0
-80.0
-90.0
-110.0
-100.0
SAT JAN 28, 2003
14:30:25
INT:REF
AMPL CAL
Video
Level
Singl
Free
Run
Trigger
Video
Level
0
1
2
3
7
6
5
4
8
9
.
-
GHz
dBm
MHz
dBmV
kHz
dBuV
BK SP
Start:99.5MHz
RBW:30 kHz
Stop:100.5MHz
VBW:100 kHz
Center:100.0MHz
Span:1.0MHz
Sweep:100.0ms
Ref
Lvl: -30.0 dBm
-30.0
-40.0
-50.0
-60.0
-70.0
-80.0
-90.0
-110.0
-100.0
SAT JAN 28, 2003
14:30:25
INT:REF
AMPL CAL
Trigger
Delay
Trigger
Freq
Free
Run
Video
Level
Singl
Trigger
Delay
Trigger
Freq
Video Level:-40 dBm
1