4.1.3. Anomalous cases during "AUTO", "R+TIMER", "R-TIMER" tests

 In case a value of Ravg or R+ or R-

higher than or equal to 5

but lower

than 99.9

was detected, at the end of

the test the instrument

emits a long

sound signal and displays one screen
similar to the screen alongside.

CAL LOW



 5.75

216

mA

5

s


AUTO

 In AUTO mode, if a Ravg or R+ or R-

higher than 99.9

was detected, at

the end of the test the instrument emits
a long sound signal and displays the
screen alongside.

CAL LOW



> 99.9

- - -

mA


AUTO

 In case the mode R+TIMER or R-

TIMER was selected and a

R+ or R-

higher than 99.9

 was detected, the

instrument emits

an intermittent

sound signal during the test, a long
sound signal at the end of the test

and displays the screen alongside.

CAL LOW



> 99.9

- - -

mA

4

s

R+


TIMER

 In case that:

R

MEASURED

-R

CALIBRATION

< 0

 the

instrument displays the screen
alongside.

CAL LOW



 0.00

219

mA

AUTO

SAVING:

The tests can be stored pressing the

SAVE key twice

(according to § 5.1).

 If the terminal voltage is higher than

10V, the instrument does not carry out
the test and displays the screen
alongside for 5 seconds after which, the
instrument displays the screen relative
to the precedent selected test mode
LOW

 under AUTO mode.

CAL LOW



 13V

VOL AG


THIS RESULT CANNOT BE SAVED.

ATTENTION: value of R

avg

higher than 5

99.9

is the

maximum value
which can be
measured in the
LOW

AUTO or

R+ or R-mode.

99.9

is the

maximum value.

AUTO mode.

AUTO mode.

ATTENTION: value of R

avg

or R+ or R-

is too

high.

ATTENTION:
value of R+ or R-
is too high.

Only in case the
mode R+TIMER
or R-TIMER was
selected.

Blinking

CAL.

ATTENTION: R

MEASURED

-R

CALIBRATION

< 0

AUTO mode.

SAVE

ATTENTION: the
test was not
effected
because
of voltage at the
terminal ends.

SAVE