SIRIUS

87

EN - 15

4.2. M

:

INSULATION RESISTANCE MEASUREMENT


The measurement is performed according to IEC/EN61557-2 and VDE 0413 part 1.

CAUTION

Before effecting the insulation test make sure that the circuit under test is not
energized and all the relative loads are disconnected.


Turn the

switch on M

 position.

Switch on the instrument.


The key

FUNC permits to select one of the following measuring modes

(which can be shown cyclically when pressing the key):

 Mode

MAN (minimum test time of 10 seconds or set by the pressing

duration on the

START/STOP key). Recommended test.

 Mode

TIMER (test duration depending on the selected interval (from 10

to 999 seconds). This test can be effected in case a minimum measuring
time is required.

FUNC