HT instruments SIRIUS87 User Manual
Page 43

SIRIUS
87
EN - 41
If the instrument detects that the earth
cable (green) is not connected, the
screen alongside is displayed for 5
seconds then the initial display is back.
Check the connections of PE conductor
under test.
LOOP
>
1999
no
PE
If a contact voltage U
t
higher than the
selected limit (U
L
) is detected the
instrument
interrupts the test and emits
a long sound signal at the end of the
test and displays the values
alongside.displays the screen
alongside.
LOOP
>
50
V
A
UL
t
0° U
L
=50V
If during the test the instrument detects
a lack of voltage the screen alongside
is displayed.
LOOP
c
P-P
- - - - - -
t
If the thermistores of the instrument are
damaged, it displays the screen
alongside.
LOOP
c
If the instrument gets overheated it
does not permit to effect the tests and
displays the message alongside. Wait
until the instrument displays the initial
screen before proceeding with
measuring.
LOOP
THE PREVIOUS RESULTS CANNOT BE SAVED.
Message
“no PE”: the
instrument does
not detect the
protection circuit.
Messagge
“hot”:
the instrument is
overheated.
P-P or P-N
or P-PE
Message
"Att
UL":
a too high
contact voltage
was detected
Message
">50" o “>25V”: the instrument detects
a contact voltage higher than the selected limit
(50V in this case).
The instrument
can display 0° or
180°
SAVE