Programming Sequence (Relay 1)

On the Con TROLL PRO, select Menu, Settings, Outputs, Relays, Relay 1.

l

Enable the relay.

l

Mode = Low Alarm

l

Parameter = A-Concentration

l

Setpoint = 4 mg/L

l

Deadband = 0.2 mg/L (This value is recommended to prevent relay chatter.)

Adjust the Setpoint and Deadband for Relay 1 to optimize aerator
performance and minimize aerator cycling.

Programming Sequence (Relay 2)

On the Con TROLL PRO, select Menu, Settings, Outputs, Relays, Relay 2.

l

Enable the relay.

l

Mode = Low Alarm

l

Parameter = A-Concentration

l

Setpoint = 3 mg/L

l

Deadband = 0.2 mg/L (This value is recommended to prevent relay chatter.)

Configuring a Dual-Parameter Light Alarm

In this scenario, the Con TROLL PRO monitors two independent tanks, each with an
RDO PRO Probe or RDO PRO-X Probe.

Probe A is placed in Tank A. The Con TROLL PRO turns on aerator A and the yellow
light when the DO falls below 4 mg/L.

Probe B is placed in Tank B. The Con TROLL PRO turns on aerator B and the red light
when the DO falls below 5 mg/L.

Relay 1 is used to implement the Tank A control function, Relay 2 is used to implement
the Tank B control function.