H) Turn on BIPOLAR CURRENT CONTROL (16) clockwise through its range, while observing
the front panel “I
O
METER” (17). The BOP output current should smoothly follow from the
maximum negative output current, over zero, to the maximum positive output current of the
BOP. Turn A-C POWER switch (1) “OFF”. Remove the short circuit from the output termi-
nals. THIS CONCLUDES THE PRELIMINARY CHECK-OUT OF THE BOP.
2.6
INSTALLATION (Refer to FIG. 2-5)
The BOP is delivered with mounted SLIDE SUPPORT BRACKETS and MOUNTING FLANGES
and ready for installation into any standard (19-inch) instrument rack. Please refer to PAR 2.4
(COOLING) if the BOP is to be mounted into a multiple rack installation.
FIGURE 2-5. RACK INSTALLATION OF THE BOP