B.15 DEBug:TREE:SEParator? QUERY
DEB:TREE:SEP?
Syntax:
Short Form: DEB:TREE:SEP?
Long Form: DEBug:TREE:SEParator?
Returns <NR1> defines the column separator
Description: Returns the number corresponding to the selected column separator when producing a text
file of the command set using DEB:TREE:DISP command.
B.16 LOCK COMMAND
LOCK
Syntax:
Short Form: LOCK <lock state>
Long Form: LOCK <lock state>
where <lock state> = OFF, EMERgency or LOCKed
Description: Controls functionality of front panel controls. When set to OFF, all front panel controls are fully
functional. When set to EMER, the front panel LOAD switch can only turn the load to the off state; all
other front panel controls except for the POWER switch are disabled. When set to LOCK, all front
panel controls except for the POWER switch are disabled.
B.17 LOCK? QUERY
LOCK?
Syntax:
Short Form: LOCK?
Long Form: LOCK?
Returns: <lock state> = OFF, EMERgency or LOCKed
Description: Indicates state of front panel controls.
B.18 SENSe:CURRent:RANGe COMMAND
SENS:CURR:RANG
Syntax:
Short Form: SENS:CURR:RANG <NR1> Long Form: SENSe:CURRent:RANGe <NR1>
Where: <NR1> = 0, 1, 2, 3 or 4
Description: Controls the placement of the decimal point during remote interface readback of current and
on the front panel AMPS display.
Display Ranges: 0 - Auto, 1 - 0.000, 2 - 00.00, 3 - 000.0, 4 - 0000.
B.19 SENSe:CURRent:RANGe? QUERY
SENS:CURR:RANG?
Syntax:
Short Form: SENS:CURR:RANG?
Long Form: SENSe:CURRent:RANGe?
Returns: <NR1> = 0, 1, 2, 3 or 4
Description: Indicates the placement of the decimal point during remote interface readback of current and
on the front panel AMPS display.
Display Ranges: 0 - Auto, 1 - 0.000, 2 - 00.00, 3 - 000.0, 4 - 0000.
B.20 SENSe:VOLTage:RANGe COMMAND
SENS:VOLT:RANG
Syntax:
Short Form: SENS:VOLT:RANG <NR1>
Long Form: SENSe:VOLTage:RANGe <NR1>
Where: <NR1> = 0, 1, 2, 3 or 4
Description: Controls the placement of the decimal point during remote interface readback of voltage, and
on the front panel VOLTS display.
Display Ranges: 0 - Auto, 1 - 0.000, 2 - 00.00, 3 - 000.0, 4 - 0000.
B.21 SENSe:VOLTage:RANGe? QUERY
SENS:VOLT:RANG?
Syntax:
Short Form: SENS:VOLT:RANG? Long
Form:
SENSe:VOLTage:RANGe?
Returns: <NR1> = 0, 1, 2, 3 or 4
Description: Indicates the placement of the decimal point during remote interface readback of voltage, and
on the front panel VOLTS display.
Display Ranges: 0 - Auto, 1 - 0.000, 2 - 00.00, 3 - 000.0, 4 - 0000.
B.22 [SOURce:]INPut[:STATe] COMMAND
INP
Syntax:
Short Form: [SOUR:]INP[:STAT] <bool>
Long Form: [SOURce:]INPut[:STATe] <bool>
where <bool> = 0 or OFF (load disengaged) or 1 or ON (load engaged)
Description: Used to engage or disengage the load from the UUT. When set to 1 the load engages at the pro-
grammed mode and corresponding setpoint value; when set to 0 the load disengages.