B.48 [source:]current:protection:state? query, B.49 [source:]current:range command, B.50 [source:]current:range? query – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 3.12 (Unit Rev 2) User Manual
Page 68: B.51 [source:]function command, B.52 [source:]function query, B.53 [source:]mode command, B.54 [source:]mode? query, B.48, Source, B.49

B-8
SERIES EL 070110
B.48 [SOURce:]CURRent:PROTection:STATe? QUERY
CURR:PROT:STAT?
Syntax:
Short Form: [SOUR:]CURR:PROT:STAT?
Long Form: [SOURce:]CURRent:PROTection:STATe?
Returns: <bool> = 1 (enabled) or 0 (disabled)
Description: Indicates whether the current protection level set using CURR:PROT is enabled (1) or disabled
(0).
B.49 [SOURce:]CURRent:RANGe COMMAND
CURR:RANG
Syntax:
Short Form: [SOUR:]CURR:RANG <NR1>
Long Form: [SOURce:]CURRent:RANGe <NR1>
where: <NR1> = 0, 1, 2, 3 or 4
Description: Controls the placement of the decimal point during remote interface readback of current, and
on the front panel AMPS display when showing current. Display Ranges: 0 - Auto, 1 - 0.000, 2 -
00.00, 3 - 000.0, 4 - 0000.
B.50 [SOURce:]CURRent:RANGe? QUERY
CURR:RANG?
Syntax:
Short Form: [SOUR:]CURR:RANG?
Long Form: [SOURce:]CURRent:RANGe?
Returns: <NR1> = 0, 1, 2, 3 or 4
Description: Indicates the placement of the decimal point during remote interface readback of current, and
on the front panel AMPS display when showing current. Display Ranges: 0 - Auto, 1 - 0.000, 2 -
00.00, 3 - 000.0, 4 - 0000.
B.51 [SOURce:]FUNCtion COMMAND
FUNC
Syntax:
Short Form: [SOUR:]FUNC <txt>
Long Form: [SOURce:]FUNCtion <txt>
where: <txt> = CURR, POW, VOLT, RES, COND or OFF
Description: Sets operating mode.
B.52 [SOURce:]FUNCtion QUERY
FUNC?
Syntax:
Short Form: [SOUR:]FUNC?>
Long Form: [SOURce:]FUNCtion?
Returns: <txt> = CURR, POW, VOLT, RES, COND or OFF
Description: Indicates operating mode.
B.53 [SOURce:]MODE COMMAND
MODE
Syntax:
Short Form: [SOUR:]MODE <txt>
Long Form: [SOURce:]MODE <txt>
where: <txt> = CURR, POW, VOLT, RES, COND, SHORT or OFF
Description: Sets operating mode: CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or OFF.
B.54 [SOURce:]MODE? QUERY
MODE?
Syntax:
Short Form: [SOUR:]MODE?
Long Form: [SOURce:]MODE?
Returns: <txt> = CURR, POW, VOLT, RES, COND, SHORT or OFF
Description: Indicates the present mode: CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or
OFF.
B.55 [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude] COMMAND
POW
Syntax:
Short Form: [SOUR:]POW[:LEV][:IMM][:AMP] <NR2>
Long Form: [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude] <NR2>
Where <NR2> = power setpoint
Description: Establishes power setpoint.