B.57 [source:]power:mode command, B.58 [source:]power:mode? query, B.59 [source:]power:protection command – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 3.12 (Unit Rev 2) User Manual
Page 69: B.60 [source:]power:protection? query, B.63 [source:]power:range command, B.56, Source, B.57, B.58, B.59

SERIES EL 070110
B-9
B.56 [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude]? QUERY
POW?
Syntax:
Short Form: [SOUR:]POW[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude]?
Returns <NR2> = power setpoint
Description: Indicates power setpoint.
B.57 [SOURce:]POWer:MODE COMMAND
POW:MODE
Syntax:
Short Form: [SOUR:]POW:MODE
Long Form: [SOURce:]POWer:MODE
Description: Selects power mode.
B.58 [SOURce:]POWer:MODE? QUERY
POW:MODE?
Syntax:
Short Form: [SOUR:]POW:MODE?
Long Form: [SOURce:]POWer:MODE?
Returns: <txt> = CURR, POW, VOLT, RES, COND, OFF
Description: Returns the present mode: CURRent, POWer, VOLTage, RESistance, CONDuctance, OFF
B.59 [SOURce:]POWer:PROTection COMMAND
POW:PROT
Syntax:
Short Form: [SOUR:]POW:PROT <NR2>
Long Form: [SOURce:]POWer:PROTection <NR2>
where <NR2> = power protection setpoint
Description: Selects power protection setpoint. This is a user-defined value which, if exceeded during operation,
causes a FAULT condition, forcing the load to disengage.
B.60 [SOURce:]POWer:PROTection? QUERY
POW:PROT?
Syntax:
Short Form: [SOUR:]POW:PROT?
Long Form: [SOURce:]POWer:PROTection?
Returns <NR2> = power protection setpoint
Description: Indicates power protection setpoint.
B.61 [SOURce:]POWer:PROTection:STATe[:LEVel] COMMAND
POW:PROT:STAT
Syntax:
Short Form: [SOUR:]POW:PROT:STAT[:LEV] <bool>
Long Form: [SOURce:]POWer:PROTection:STATe[:LEVel] <bool>
where: <bool> = 1 (enabled) or 0 (disabled)
Description: Clears or sets power protection trip.
B.62 [SOURce:]POWer:PROTection:STATe[:LEVel]? QUERY
POW:PROT:STAT?
Syntax:
Short Form: [SOUR:]POW:PROT:STAT[:LEV]?
Long Form: [SOURce:]POWer:PROTection:STATe[:LEVel]?
where: <bool> = 1 (enabled) or 0 (disabled)
Description: Indicates whether the power protection level set using POW:PROT is enabled (1) or disabled
(0).
B.63 [SOURce:]POWer:RANGe COMMAND
POW:RANG
Syntax:
Short Form: [SOURce:]POW:RANGe <NR1>
Long Form: [SOURce:]POWer:RANGe <NR1>
where: <NR1> = 0, 1, 2, 3 or 4
Description: Controls the placement of the decimal point during remote interface readback of power, and on
the front panel KW display when showing power. Display Ranges: 0 - Auto, 1 - 0.000, 2 - 00.00, 3
- 000.0, 4 - 0000.