Start-up – KROHNE OPTISONIC 6300 Quickstart EN User Manual

Page 43

Advertising
background image

START-UP

3

43

OPTISONIC 6300

www.krohne.com

07/2009 - 7.30959.24.00 - QS OPTISONIC 6300 R04 en

• Go through menu X1...X7 as described in section "Installation menu" in chapter "General

instructions for programming". Correct X5 if needed.

• X9.1: Press enter
• X9.2: Press enter
• X9.3: Press enter
• X9.4: Press enter
• X9.5: Read the preliminary volume flow. Press enter
• X9.6: Check signal

Figure 3-13: Construction (field version)

1 Cover, electronics compartment
2 Cover, terminal compartment for power supply and inputs/outputs
3 Cable entry for power
4 Cable entry for inputs/outputs
5 Cable entry for sensor cable
6 Cover, sensor terminal compartment

INFORMATION!
See also the section "Installation menu" in chapter "General instructions for programming".

CAUTION!
Advice on signal strength:

Advice on signal strength:

Advice on signal strength:

Advice on signal strength:
Signal > 75%:

Signal > 75%:

Signal > 75%:

Signal > 75%: good signal, optimization loop not needed
Signal 50...75%:

Signal 50...75%:

Signal 50...75%:

Signal 50...75%: fairly good signal, optimization loop can improve the signal
Signal 10...50%:

Signal 10...50%:

Signal 10...50%:

Signal 10...50%: low signal, optimization loop needed
Signal < 10%:

Signal < 10%:

Signal < 10%:

Signal < 10%: bad or no signal, check settings in menu X6, increase transducer distance and/or
go into the optimization loop.