Displaying units
You can select the parameter units.
for pH:
pH
or
mV
for Redox
mV
or
Volt
for ISE
connectn. or
Atom
(NH
4
NH
4
-N)
"Display – Units"
Auto clean function
If necessary, set the rinse function for immersion holders here. Our recommendation:
Rinse medium Air
Interval: 0,5 hr.
Rinse duration: 4 sec.
Hold time:
30 sec..
Automatic:
yes
Rinse medium water
Interval: 2,0 hr.
Rinse duration: 20 sec.
Hold time:
30 sec.
Automatic:
yes
"Rinse – Interval / Duration / Hold time / Automatic"
CAUTION!
If you use water as the rinse medium, it is essential that you activate the winter service in addition
(against risk of frost)!