KROHNE OPTISENS PAM 2080 EN User Manual

Page 28

Advertising
background image

www.krohne.com

28

Displaying units

You can select the parameter units.

for pH:

pH

or

mV

for Redox

mV

or

Volt

for ISE

connectn. or

Atom

(NH

4

NH

4

-N)


"Display – Units"

Auto clean function

If necessary, set the rinse function for immersion holders here. Our recommendation:

Rinse medium Air

Interval: 0,5 hr.

Rinse duration: 4 sec.

Hold time:

30 sec..

Automatic:

yes

Rinse medium water

Interval: 2,0 hr.

Rinse duration: 20 sec.

Hold time:

30 sec.

Automatic:

yes


"Rinse – Interval / Duration / Hold time / Automatic"

CAUTION!

If you use water as the rinse medium, it is essential that you activate the winter service in addition
(against risk of frost)!