Installation – KROHNE OPTISENS AAS 2000 EN User Manual
Page 14
3
INSTALLATION
14
OPTISENS AAS 2000
www.krohne.com
12/2008 • MA AAS 2000 R01 en
Removing an old electrode (see figure above)
•
Make sure automatic flushing is disabled. This can be checked at the MAC 080 converter in the
Cleaning
Cleaning
Cleaning
Cleaning menu of the AAS 2000 sensor ( on page 22).
•
If using the protection plate, pull it off before changing the electrode 1.
•
Loosen the flushing head by turning it counterclockwise 2.
•
Gently pull the flushing head out of the sensor housing 3
•
Unplug the electrode 4.
•
Remove the electrode from the flushing head by pressing firmly on the backside of the
electrode near the cable connector 5.
•
Ensure that the O-ring that seals the flushing head is completely seated in the respective
groove in the sensor body.
•
If a new electrode is to be mounted, please follow the previous instructions ("Installing a new
electrode").
Figure 3-4:
Removing an old electrode
CAUTION!
Make sure the electrode housing is dry before removing the flushing head. Water may come in to
the housing when the flushing head is removed.