Operation – KROHNE OPTISENS AAS 2000 EN User Manual

Page 26

Advertising
background image

5

OPERATION

26

OPTISENS AAS 2000

www.krohne.com

12/2008 • MA AAS 2000 R01 en

Performing an air calibration

•

Fill a bucket halfway with water.

•

Place the sensor about 30...45 cm/1...1.5 ft above the water surface to create a humid

environment.

•

Cover the bucket with a towel.

•

Select the sensor to be calibrated in the menu by using ↑ or ↓.

•

Press ^ for approximately five seconds to enter the sensor menu.

•

Select Calibrate > Airpres.mB

Calibrate > Airpres.mB

Calibrate > Airpres.mB

Calibrate > Airpres.mB and press ^.

•

Enter the current air pressure in mbar using ↑ or ↓ to change a digit and ^ to move on to the

next digit. It is sufficient to specify the pressure within 10 mbar (see table below for
conversion from inHg to mbar).

•

Select Calibrate > Calibrate

Calibrate > Calibrate

Calibrate > Calibrate

Calibrate > Calibrate and press Press ^.

•

Select Air

Air

Air

Air using ↑ or ↓ (three choices can be made, "Air"/"No"/"Zero").

•

Press ^.

•

Wait for the sensor to be calibrated. A successful calibration will take anywhere between 5...15

minutes for the sensor to stabilize.

i

A dialog box saying "Calibration done" will come up after successful calibration.

•

Press ^ to return to the previous menu.

•

Select Calibrate > Calibrated

Calibrate > Calibrated

Calibrate > Calibrated

Calibrate > Calibrated and press Press ^.

•

Enter the date of the air calibration using ↑ or ↓ to change a digit and ^ to move on to the next

digit.

•

Press ^ to return to the previous menu.

i

The sensor is air calibrated.

The following appear in the calibration menu as read-only:

•

Temp (info)

Temp (info)

Temp (info)

Temp (info): Displays the current temperature of the sensor.

•

Test (info)

Test (info)

Test (info)

Test (info): Displays the current measured dissolved oxygen value. Immediately after
calibration and before the sensor is placed in the liquid, this value should closely resemble
Setvalue. A reading outside of an acceptable range may indicate that the electrode needs to
be replaced.

•

Setvalue

Setvalue

Setvalue

Setvalue: Ideal value calculated using temperature and air pressure.