The pH sensors are available with different diaphragms:
pH sensor with open diaphragm and Pt100
open diaphragm (microscopically small hole in the glass); integrated Pt100 temperature
measurement; suitable for cooling water, waste water and chemical industry applications.
pH sensor with ceramic diaphragm
Ceramic diaphragm; suitable for drinking water and pool industry applications.
pH sensor with PTFE diaphragm
Dirt-repellent Teflon diaphragm; suitable for waste water, surface water and process water
applications.
2.3 Nameplate
The sensor type is specified on the labelling of the sensor package and on the sensor itself.
Figure 2-3: Diaphragm types
1 Reference electrode
2 KCl solution
3 Inner pH 7 buffer solution
4 Measuring electrode
5 PTFE diaphragm
6 Ceramic diaphragm
7 open diaphragm
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.