TECHNICAL DATA

42

OPTISENS PH 9500

www.krohne.com

08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en

The voltage change of a pH sensor at 25°C / 77°F is around 59 mV for each pH unit. This is also
called the slope of the pH sensor. The slope is temperature dependent and decreases over life
time of the sensor.

To compensate for the temperature dependency of the pH measurement, the temperature of the
medium can be measured and automatically compensated in the signal converter.

Figure 6-2: Optimal slope at 25°C / 77°F

Figure 6-3: Temperature dependency of the slope