2 ph temperature dependency, Appendix – KROHNE OPTISENS PH 9500 User Manual
Page 47

APPENDIX
7
47
OPTISENS PH 9500
www.krohne.com
08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en
7.2 pH temperature dependency
The output from a pH-sensor varies with the temperature in a predictable way. The size of the
variation depends on both the temperature and the pH being measured.
At pH 7 and 25°C / 77°F the temperature error is zero. If temperature or pH changes the
temperature error is calculated using the following formula:
0.03 pH-difference / pH or 0.03 pH-difference / K
°
C
°
F
pH
pH
pH
pH
pH
pH
pH
pH
pH
pH
pH
pH
5
41
2.30
3.24
4.18
5.12
6.06
7.00
8.06
9.12
10.18 11.24 12.30 13.36
15
59
2.15
3.12
4.09
5.06
6.03
7.00
8.03
9.06
10.09 11.12 12.15 13.18
25
77
2.00
3.00
4.00
5.00
6.00
7.00
8.00
9.00
10.00 11.00 12.00 13.00
35
95
1.85
2.88
3.91
4.94
5.97
7.00
7.97
8.94
9.91
10.88 11.85 12.82
45
113
1.70
2.76
3.82
4.88
5.94
7.00
7.94
8.88
9.82
10.76 11.70 12.64
55
131
1.55
2.64
3.73
4.82
5.91
7.00
7.91
88.2
9.73
10.64 11.55 12.46
65
149
1.40
2.52
3.64
4.76
5.88
7.00
7.88
8.76
9.64
10.52 11.40 12.28
75
167
1.25
2.40
3.55
4.70
5.85
7.00
7.85
8.70
9.55
10.40 11.25 12.10
85
185
1.10
2.28
3.46
4.64
5.82
7.00
7.82
8.64
9.46
10.28 11.10 11.92
95
203
0.95
2.16
3.37
4.58
5.79
7.00
7.79
8.58
9.37
10.16 10.95 11.74