2 mounting to a flow-through holder, Installation – KROHNE OPTISENS PH 9500 User Manual

Page 18

Advertising
background image

3

INSTALLATION

18

OPTISENS PH 9500

www.krohne.com

08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en

3.7.2 Mounting to a flow-through holder

Installing a new sensor

• Make sure that the O-ring 5 and the washer 4 on the sensor are assembled in the sequence

indicated in the drawing.

• Screw the sensor into the female thread 2 of the flow-through holder 1. Tighten the sensor

by hand.

• If you have not yet established the electrical connection to the signal converter, leave the

protective cap on the sensor until you establish the electrical connection.

WARNING!
Ensure that the pipe is without pressure before installing or removing a sensor!

INFORMATION!
The flow-through holder is an optional accessory and not part of the standard scope of delivery.
It has to be installed horizontally in pump or sample lines or directly in the process.

Figure 3-8: Possible mounting positions of the flow-through holder

1 Mounting in an outlet pipe
2 Mounting in a bypass pipe
3 Valve
4 Flow-through holder

Figure 3-9: Installing the sensor into the flow-through holder

1 Flow-through holder
2 Female thread
3 Sensor thread
4 Washer
5 O-ring
6 Process connection
7 Flow direction
8 Protective cage