Technical data – KROHNE IFC 050 Converter EN User Manual
Page 75

TECHNICAL DATA
8
75
IFC 050
www.krohne.com
08/2013 - 4002184002 - MA IFC 050 R02 en
Current output
Current output
Current output
Current output
Output data
Flow
Settings
Without HART
Without HART
Without HART
Without HART
®
Q = 0%: 0…20 mA; Q = 100%: 10…21.5 mA
Error identification: 20…22 mA
With HART
With HART
With HART
With HART
®
Q = 0%: 4…20 mA; Q = 100%: 10…21.5 mA
Error identification: 3…22 mA
Operating data
Operating data
Operating data
Operating data
Basic I/Os
Basic I/Os
Basic I/Os
Basic I/Os
Active
Observe connection polarity.
U
int, nom
= 20 VDC
I ≤ 22 mA
R
L
≤ 750 Ω
HART
®
at terminals A
Passive
Observe connection polarity.
U
ext
≤ 32 VDC
I ≤ 22 mA
U
0
≤ 2 V at I = 22 mA
R
L, max
= (U
ext
- U
0
) / I
max
HART
®
at terminals A
HART
HART
HART
HART
®
Description
HART
®
protocol via active and passive current output
HART
®
version: V5
Universal Common Practice HART
®
parameter: completely supported
Load
≥ 250 Ω at HART
®
test point;
Note maximum load for current output!
Multi-drop mode
Yes, current output = 4 mA
Multi-drop address adjustable in operation menu 1…15