DESCRIPTION OF HART INTERFACE

9

113

IFC 100

www.krohne.com

07/2010 - 4000041004 - MA IFC 100 R04 en

Menu View

Display

volume flow / mass flow / flow speed / conductivity / coil temperature / diagnosis value / device

status

Counter

counter 1

Opt

/ counter 2

Opt

Yt diagram

volume flow / mass flow

Outputs

current output A

Opt

measured value

Opt

/

A

Opt

/

% range A

Opt

frequency output D

Opt

measured value

Opt

/

D

Opt

/

% range D

Opt

Device Status

device

C number

Rd

/ device serial no.

Rd

/ electronic serial no.

Rd

HART

tag / manufacturer

Rd

/ write protect

Rd

/

model

Rd

/ device ID / universal rev.

Rd

/

device rev.

Rd

/ software rev.

Rd

/ hardware rev.

Rd

/

date

Rd

/ final assembly no.

Rd

/ sensor serial no.

Rd

Standard (overview)

Primary variable out of limits

Non-primary variable out of limits

Primary variable analogue output saturated

Primary variable analogue output fixed

Cold start

Configuration changed

Field device malfunction

Failure (device)

F error in device / F IO1 / F parameter / F configuration /

F display / F sensor electronic / F sensor global / F sensor local /

F field current local / F current output A / F software user interface /

F hardware settings / F hardware detection /

F RAM/ROM error IO1

Failure (application)

F application error / F empty pipe / F flow rate too high /

F field frequency too high / F DC offset / F open circuit A /

F over range A (current) / F over range D (pulse) /

F active settings / F factory settings /

F backup 1 settings / F backup 2 settings

Out of specification

S out of specification / S pipe not full / S pipe empty /

S linearity / S flow profile / S electrode noise /

S gain error / S electrode symmetry / S field coil broken /

S field coil bridged / S field current deviation /

S field frequency too high / S electronic temperature /

S coil temperature / S overflow counter 1 / S overflow counter 2 /

S Backplane Invalid

check request

C checks in progress / C test sensor

information

I counter 1 stopped / I counter 2 stopped / I power fail /

I over range display 1 / I over range display 2 /

I backplane sensor / I backplane settings /

I backplane difference / I optical interface

Circuit board info

Toolbar

Status Bar

Update