Electrical connections – KROHNE IFC 100 Converter EN User Manual

Page 42

Advertising
background image

4

ELECTRICAL CONNECTIONS

42

IFC 100

www.krohne.com

07/2010 - 4000041004 - MA IFC 100 R04 en

Pulse/frequency output passive

• U

ext

≤ 32 VDC

• f

max

in operating menu set to 100 Hz < f

max

≤ 10 kHz:

I ≤ 100 mA at ≤ 10 kHz (overcontrol up to f

max

≤ 12 kHz)

open:
I ≤ 0.1 mA at U

ext

= 5 V

I ≤ 0.5 mA at U

ext

= 24 V

I ≤ 0.7 mA at U

ext

= 32 V

closed:
U

0, max

= 0.8 V at I ≤ 1 mA

U

0, max

= 1.5 V at I ≤ 10 mA

U

0, max

= 3.5 V at I ≤ 100 mA

• If the following maximum load resistance R

L, max

is exceeded, the load resistance R

L

must be

reduced accordingly by parallel connection of R:
f ≤ 1 kHz: R

L, max

= 10 kΩ

f ≤ 10 kHz: R

L, max

= 2 kΩ

• The minimum load resistance R

L, min

is calculated as follows:

R

L, min

= (U

ext

- U

0

) / I

max

• Can also be set as status output; for the electrical connection refer to status output

connection diagram.

• The output is closed when the device is de-energized.

INFORMATION!
•

For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from
electrical interferences (EMC).

•

Shielding takes place at the electrical connection (S) of the output terminal block.

CAUTION!
Observe connection polarity.

Figure 4-18: Pulse/frequency output passive P

p

.book Page 42 Friday, July 9, 2010 12:29 PM