SERVICE
7
57
H250
www.krohne.com
04/2011 - 4000269302 MA H250-R02 en
The brake with the retainer ring can be clipped onto the pointer cylinder independently of the
built-in components (ESK2A, limit switch, counter). When installing the brake, note that the slit
between the brake magnets is only about 3 mm and the material thickness of the aluminium
pointer vane is 1mm.
• Clip on Eddy current brake 1.
• Turn brake slightly in a clockwise direction 2.
• Check that the pointer vane can be moved between the magnets without touching them.
• Screw in turnbuckle 2.
7.2.4 Retrofitting limit switch
• Remove counter module (if available).
• Loosen the locking screw 2 on the contact pointer.
• Merge contact pointer 1 in the middle.
• Insert the contact module into the third slot of the bracket until the semi-circle 3 surrounds
the pointer cylinder.
The contact module connecting terminals feature a pluggable design and can be removed in
order to connect the cables.
MA_H250_R02_en_269302_PRT.book Page 57 Tuesday, April 12, 2011 10:17 AM