2 ambient / process temperatures – KROHNE OPTIMASS with MFC 300 EN User Manual

Page 62

Advertising
background image

62

OPTIMASS

8

OPTIMASS 8000/9000

8.1 Specific Installation Guidelines

•

Tighten flange bolts evenly.

•

Do not stress the sensor mechanically. Clamp and support the connecting pipework accord-
ingly.

•

It is permissible to support the weight of the meter on the square body.

•

Cavitation and mechanical vibration should be avoided.

•

Use of standard pipework reducers at the flange is allowed. Avoid extreme changes in pipe
size (step changes).

•

DO NOT use flexible hoses to connect directly to the meter.

•

Where the ambient temperature is below 0°C, the meter should be mounted vertically, (or
horizontally with converter upwards) in order to prevent freezing or condensation in the
housing.

Horizontal Mounting:

1

For measuring liquids, install so that the measuring tube faces downwards. This prevents a
build up of gasses when there is no liquid flowing.

2

For measuring gasses, install so that the measuring tube faces upwards. This prevents a
build up of liquid when there is no gas flowing.

8.2 Ambient / Process Temperatures

The specified and approved ambient and process temperatures must be observed.

Note:

For additional temperature limits in hazardous area applications, reference should be made to
the publication “

Guidelines for the use of Coriolis Meters in Hazardous Areas”.

Where meters are mounted in direct sunlight, it is recommended to install a sunshade. This is
particularly important in countries with high ambient temperatures.

The maximum differential temperature between the process and ambient temperature without
insulation is 80°C or 176°F

8000

9000

°C

°F

°C

°F

Process

Safe Area

-180...+230

-292...+446

0... +350

32...662

ATEX/FM/CSA

Compact

-40...+190

-40...+374

-

-

ATEX/FM/CSA

Remote

-40...+230

-40...+446

0...350

32...662

Ambient

Compact Aluminium

-40...+60

-40...+140

1

-

-

Compact Stainless

-40...+55

-40...+130

-

-

Remote

-40...+65

-40...+149

-40...+65

-40...+149

1

Temp range: -40...+65°C (-40...+149°F) on certain I/O options. Pleae call
Krohne for details

1

2

Advertising