5 mains power connection versions c, f and w – KROHNE OPTIMASS with MFC 300 EN User Manual

Page 75

Advertising
background image

75

0

Cover, electronics compartment

1

Locking screw, 1/2 turn left/right to open/close cover (2)

2

Cover for the three separate terminal compartments for power, sensor connection and
inputs/outputs

3

Safety lever to open cover (1)

4

Sensor terminal compartment, open separate cover

5

Terminal compartment, outputs/inputs

6

Power terminal compartment, open separate shock-hazard protection cover

7

Entry fitting for sensor cable

8

Two cable entries for outputs/inputs

9

Cable entry for power supply

9.5 Mains Power Connection versions C, F and W

PLEASE NOTE!

•

The degree of protection in relation to IP 65 and 67 to IEC 529 / EN 60529, (NEMA 4 / 4X) is
dependent on the version.

•

The housings of the flowmeters, which are designed to protect the electronic equipment
from dust and moisture, should be kept closed at all times. Creepage distances and clear-
ances are dimensioned to VDE 0110 and IEC 664 for pollution severity 2. Supply circuits are
designed for overvoltage category III and the output circuits for overvoltage category II

•

Fuse protection for the infeed power circuit, and also a disconnecting device (switch, circuit
breaker) to isolate the signal converter must be provided.

100...230 Volt AC (tolerance range: -15%...+10%)

•

Note the data on the nameplate, power supply voltage and frequency range (50...60 Hz).

•

The protective ground conductor PE of the power supply must be connected to the separate
terminal in the terminal compartment of the signal converter.

•

Connection diagrams I - II for the power supply and the electrical connection between flow
sensor (primary head) and signal converter are provided in this section.

12...24 Volt DC (tolerance range: -25%...+30%)

•

Note the data on the instrument nameplate!

•

For reasons to do with the measurement process, connect a functional ground FE to the
separate U-clamp terminal in the terminal compartment of the signal converter.

•

When connecting to functional extra-low voltages, provide a facility for protective separation
(PELV) (VDE 0100 / VDE 0106 and/or IEC 364 / IEC 536 or relevant national regulations).

•

Connection diagrams I - II for the power supply and the electrical connection between flow
sensor and signal converter converter are provided in this section.

1

2

0

3

4

5

6

7

8

9