INSTALLATION
3
25
OPTIBAR PC 5060 C
www.krohne.com
04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
3.12 Measurement setup for liquid measurement
The following points should be observed in this application:
• The signal converter must be mounted below the measuring point.
3.13 Measurement setup for level measurement
The following points should be observed in this application:
• The pressure transmitter should be mounted below the lowest level.
• The pressure transmitter should be protected from filling/emptying current and agitator
surges when mounted.
Figure 3-9: Measurement setup for liquid measurement
1 Tapping point
2 Shut-off valve
3 Pressure transmitter
Figure 3-10: Measurement setup for level measurement
1 Pressure transmitter
2 Tank