Technical data – KROHNE OPTIBAR PC 5060 C EN User Manual

Page 64

Advertising
background image

7

TECHNICAL DATA

64

OPTIBAR PC 5060 C

www.krohne.com

04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en

Standard length

5 m / 16.40 ft

Max. length

25 m / 82.02 ft

Min. bending radius

25 mm / 0.98" at 25°C / 77°F

Diameter

approx. 8 mm / 0.31"

Colour

Non-Ex version: black

Ex version: blue

Mechanical - IP 68 (25 bar)

Mechanical - IP 68 (25 bar)

Mechanical - IP 68 (25 bar)

Mechanical - IP 68 (25 bar)

Connecting cable between IP 68 device and external housing

Construction

Four wires, one pressure compensation capillary, one suspension strand, braid, metal foil

and casing

Core cross-section

0.5 mm

2

/ AWG 20

Wire resistance

< 0.036 Ω/m / 0.011 Ω/ft

Standard length

5 m / 16.40 ft

Max. length

25 m / 82.02 ft

Min. bending radius

25 mm / 0.98" at 25°C / 77°F

Diameter

approx. 8 mm / 0.31"

Colour

Blue

Cable entry / Connector

External housing

1 x cable gland M20 x 1.5 (cable Ø 5...9 mm), 1 x blind plug M20 x 1.5 or

1 x connector (depending on version), 1 x blind plug M20 x 1.5

Spring load terminals for

core cross-section to

2.5 mm

2

/ AWG 14

Electrical

Electrical

Electrical

Electrical

Supply voltage

Non-Ex device: 9.6...35 VDC

Ex ia device: 9.6...30 VDC

Ex d device: 9.6...35 VDC

Ex ia d device: 15...35 VDC

Reverse polarity

protection

Integrated

Permissible residual ripple

Non-Ex devices

for U

n

12 VDC (9.6 < UB < 14 VDC) ≤0.7 V

eff

(16...400 Hz)

Ex ia d devices and

Ex ia devices

for U

n

24 VDC (18 < UB < 35 VDC) ≤1.0 V

eff

(16...400 Hz)

Load

R

L, max

=(UB-9.6) / 22 mA

Potential connection in

device

Electronics: not electrically isolated

Ground terminal: galvanically connected with process connection

Overvoltage category

III

Protection class

II

.book Page 64 Friday, May 2, 2014 12:49 PM

Advertising