Start-up – KROHNE OPTIBAR PC 5060 C EN User Manual

Page 50

Advertising
background image

5

START-UP

50

OPTIBAR PC 5060 C

www.krohne.com

04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en

5.8.1 Error codes

Failure

Check function

Code
Text message

Cause

Action or description

F013

No measurement value available

No valid measured value

available

Overpressure, underpressure,

measuring cell defective

F017

Adjustment span too small

Adjustment not within

specification

Change adjustment according to

the limit values

F025

Error in the linearisation table

Index markers are not

continuously rising, for example

illogical value pairs

Check linearisation table,

Delete table/Create new

F36

No operable sensor software

Failed or interrupted software

update

Repeat software update

Check electronics version

Exchange the electronics or send

device for repair

F40

Error in the electronics

Hardware defect

Exchange the electronics or send

device for repair

F041

Error in the electronics

No connection to sensor

electronics

Check connection to sensor

electronics (with remote version)

F080

General software error

Disconnect operating voltage

briefly

F113

Communication error with the

display and adjustment module,
operating software PACTware

TM

disturbed

EMC interference

Remove EMC influences

F125

Inadmissible electronics

temperature

Temperature of the electronics in

the non-specified section

Check ambient temperature

Isolate electronics

Use device with higher

temperature range

F260

Error in the calibration

Error in the calibration carried

out in the factory

Error in EEPROM

Exchange the electronics

Send device for repair

F261

Error in the configuration

Error during setup

Error when carrying out a reset

Repeat setup,

Repeat reset

F265

Measurement function disturbed

Sensor no longer carries out a

measurement

Carry out a reset

Disconnect operating voltage

briefly

F266

Inadmissible operating voltage

Operating voltage is below the

specified range

Check electrical connection - if

necessary, increase operating

voltage

Code
Text message

Cause

Action or description

C700

Simulation active

A simulation is active

Finish simulation

Automatic end after 60 minutes

.book Page 50 Friday, May 2, 2014 12:49 PM