KROHNE OPTISONIC 7060 EN User Manual

Page 21

Advertising
background image

OPTISONIC 7060

Electrical & mechanical installation manual 730962.31.00 page 21 of 52

Typical installation configuration:

Fig. 3.2: OPTISONIC 7060 installation in the pipeline for unidirectional use

1. Temperature

measurement point


The choice of the installation configuration depends on type and extent of the flow

disturbance at the installation position (according to TR G13).

Type of disturbance

Possible installation configuration

None
Elbow, reducer
Double elbow out of plane, T piece

Configuration 1 (fig. 3.2)

Gas pressure controller with/ without muffler
Diffuser
Diffuser with swirling flow

Configuration 1 (fig. 3.2), >= 20 DN distance
from device and not “in sight” of the flowmeter


For bidirectional use, inlet and outlet section at both sides of the meter body shall be of

identical, independent of the configuration used (see Fig. 3.3). The temperature

measuring point shall in this case be installed at a distance of 5x DN to the meter body

(configuration

1),


Fig. 3.3: OPTISONIC 7060 installation in the pipeline for bidirectional use

1. Temperature

measurement point

3.3 Mechanical

Installation

Work on the pipelines in preparation to installing the gas flow meter is not included in the

scope of delivery.


It is recommended to use the following tools and auxiliaries for proper installation of the

OPTISONIC

7060:

• Hoisting gear or fork lift (load capacity according to the weight information given on the

type

plate)

• Suitably sized ring wrench for flange mounting

• Sealing and separating agents

• Bolt

lubricant

• Leak detection spray