KROHNE OPTISONIC 7060 EN User Manual

Page 26

Advertising
background image

OPTISONIC 7060

Electrical & mechanical installation manual 730962.31.00 page 26 of 52

3.4.3

Checking the cable loops


To verify the cables to be correctly connected, check the cable loops. Proceed as follows:

• Disconnect both ends of the cable of the loop to be tested.

• This is to prevent connected devices from interfering with the measurement.

• Test the entire cable loop between CONVERTER and final device by measuring the loop

resistance.

• If you want to test the insulation resistance as well, you must disconnect the cables at

the electronics unit before using the insulation resistance tester.


Important


The test voltage applied would seriously damage the electronics unit!

• Reconnect all cables after the loop resistance test.


Important

• In non-intrinsically safe installations, only open the terminal boxes protected for use in

hazardous areas and connect/ disconnect cables when the system is disconnected from

the power supply.

• The front housing cover (with viewing panel) must only be opened when the system is

disconnected from the power supply and no less than 10 minutes after the system has

been

switched

off.

• Incorrect cabling may cause the OPTISONIC 7060 to fail.

This will lead to the annulment of any warranty claims. The manufacturer excludes any

liability for consequential damage.

IMPORTANT

IMPORTANT