Viewing sensor signal conditions – Diagnose Fct. 2.9

Menu 2.9 allows the viewing of eight parameters

Display

Key

Line 1

Line 2

Fct. 2.9

Diagnose

↑

Fct. 2.9.1

This displays the temperature of the measuring tube.

By pressing the → key, the temperature will be indicated.

By pressing the ↵ key, you will be returned to the Function display.

↑

Fct. 2.9.2

Strain M. T.

This displays the strain value of the measuring tube stain gauge in ohms (where fitted).

↑

Fct. 2.9.3

Strain I. C.

This displays the strain value of the inner cylinder strain gauge in ohms (where fitted).

↑

Fct. 2.9.4

Tube Freq.

This displays the current resonant frequency of the sensor. This value is primarily used to

calculate the density of the process fluid.

↑

Fct. 2.9.5

Drive energy

This value indicates the current to the driver in percent. The heavier the fluid, the higher the

number. Air entrainment also shows up as a higher number.

↑

Fct. 2.9.6

Sensor A

↑

Fct. 2.9.7

Sensor B

These display the sensor signal level. Should be as per Menu 5.2.4, and be within 2% of

each other.

↑

Fct. 2.9.8

Comm.Errors

This displays the number of serial communication errors between the Front End (sensor

mounted) and the converter electronics since power up. Normally displays 0


Viewing hardware and software versions fitted Fct. 2.10

Menu 2.10 allows the viewing of the hardware and software versions installed in the mass flowmeter.

Display

Key

Line 1

Line 2

Fct.

2.10

Versions

→

Fct. 2.10.1

Backend SW

This displays the software version in the converter MFC 050/051

→

Fct. 2.10.2

Backend HW

This displays the hardware version of the converter MFC 050/051

→

Fct. 2.10.3

Frontend SW

This displays the software version installed in the front end electronics mounted on the

sensor.