Martel Electronics BetaGauge 301 User Manual

Page 28

Advertising
background image

10.5 Remote Commands and Error Codes

The following tables list all commands, and their descriptions, that are accepted by the
calibrator.

Table 5: Common Commands

Command Description

*CLS

(Clear status.) Clears the error queue.

*IDN?

Identification query. Returns the manufacturer, model number, and firmware
revision level of the Calibrator.

*RST

Resets the calibrator to the power up state.

Table 6: Calibrator Commands

Command Description

DAMP

Turns Damp on or off.

DAMP?

Returns if DAMP is on/off

DISPLAY

Turns on/off the displays specified in the command

DISPLAY?

Returns which displays are on/off

FAULT?

Returns the most recent error code

ERROR_LOOP

Turns loop power on or off in percent error mode

ERROR_LOOP?

Returns the current state of loop power in error mode

ERROR_MODE

Turns percent error mode on or off

ERROR_MODE?

Returns whether percent error mode is on or off

ERROR_PORT

Set the pressure port for percent error mode

ERROR_PORT?

Returns the pressure port for percent error mode

FUNC

Sets the display mode as specified in the command

FUNC?

Returns the current mode of the upper, middle, and lower display

HI_ERR

Sets the 100% of span limit for percent error mode

HI_ERR?

Returns the 100% of span limit for percent error mode

LOCAL

Returns user to manual operation of the calibrator

LOCKOUT

Locks out the keypad of the calibrator in remote operation

LO_ERR

Sets the 0% of span limit for percent error mode

LO_ERR

Returns the 0% of span limit for percent error mode

PRES_UNIT

Set the pressure unit for the indicated display

PRES_UNIT?

Returns the pressure from the indicated display

REMOTE

Puts the calibrator in remote mode

ST_START

Starts a switch test

ST_OPEN?

Returns pressure value at which the switch opened

ST_CLOSE?

Returns pressure value at which the switch closed

ST_DEAD?

Returns pressure value of the deadband of the switch

26