To set loop power on send ERROR_LOOP ON.

ERROR _LOOP?

Returns the current state of loop power in percent error mode.

For example:

If you send ERROR_LOOP? It will return ON if loop power is on in error mode.

ERROR_ MODE

Turns percent error mode on and off.

For example:

To turn on percent error mode send ERROR_MODE ON.

ERROR _ MODE?

Returns the current state of percent error mode.

For example:

If you send ERROR_MODE? It will return ON if the calibrator is in percent error
mode.

ERROR_ PORT

Sets the pressure port for percent error.

For example:

To set the pressure port for percent error to [P1] send ERROR_ PORT P1.

ERROR _ PORT?

Returns the current pressure port for percent error mode.

For example:

If you send ERROR _PORT?, it will return P1 if the pressure port in percent
error is [P1].

FUNC

Sets the display indicated in argument one to the function indicated in argument 2.

For example:

To set the lower display to P1 Pressure mode send FUNC LOWER,P1.

FUNC?

Returns the current mode of all displays. For example if the calibrator is set to [P1]
ST on the upper display, [P1] on the middle, and the lower display is off, FUNC?
Would return:

30