Table 8: Error Codes

Error Number

Error Description

100

A non-numeric entry was received where it should be a numeric entry

101

Too many digits entered

102

Invalid units or parameter value received

103

Entry is above the upper limit of the allowable range

104

Entry is below the lower limit of the allowable range

105

A required command parameter was missing

106

An invalid command parameter was received

107

Pressure not selected

108

Invalid sensor type

109

Pressure module not connected

110

An unknown command was received

111

Bad Parameter received

112

The serial input buffer overflowed

113

Too many entries in the command line

114

The serial output buffer overflowed

10.6 Entering Commands

Commands for the calibrator may be entered in upper or lower case. There is at least
one space required between the command and parameter, all other spaces are optional.
Almost all commands for the calibrator are sequential; any overlapped commands will be
indicated as such. This section will briefly explain each of the commands and describe
their general use, which will include any parameters that may be entered with the
command as well as what the output of the command is.

10.6.1 Common Commands

*CLS

Clears the error queue. Also terminates all pending operations. When writing programs,
use before each procedure to avoid buffer overflow.

*IDN?

Returns the manufacturer, model number, and firmware revision of the Calibrator. For
example: