Remote operation – Martel Electronics BetaGauge 321A User Manual

Page 28

Advertising
background image

26

12. Remote Operation

12.1 Remote Interface

The calibrator can be remotely controlled using a PC terminal, or by a computer program
running the calibrator in an automated system. It uses an RS-232 serial port connection
for remote operation. NOTE: To use the remote control option a custom RS-232
cable must be purchased from Martel (LEM232). To contact Martel refer to
Section 1.1 of this manual. With this connection the user can write programs on the
PC, with Windows languages like Visual Basic to operate the calibrator, or use a Windows
terminal, such as Hyper Terminal, to enter single commands. Typical RS-232 remote
configurations are shown in Figure 13.

Figure 13. Calibrator-to-Computer Connection

12.2 Setting up the RS-232 Port for Remote Control

Note: The RS-232 connection cable should not exceed 15m unless the load capacitance
measured at connection points is less than 2500pF.

Serial parameter values:

9600 baud

8 data bits

1 stop bit

no parity

Xon/Xoff

EOL (End of Line) character or CR (Carriage Return) or both

The LEM232 cable is used for RS-232 communications from the calibrator to a computer.
If the computer only has USB type ports, a USB to RS-232 converter will be needed.
These can be obtained from most office supply and computer stores. To connect the
calibrator to the computer, attach the LEMO connector end of the cable to the pressure

Advertising
This manual is related to the following products: