CPRT_COEFB

This command is used for entering a custom RTD into the calibrator. The numeric
value entered after the command will be set as the second coefficient of the
polynomial used by the custom RTD.

For example:

CPRT_COEFB –5.774999E-07 enters –5.774999E-07 as coefficient B.

CPRT_COEFB?

Returns the number, which was entered for the second coefficient of the polynomial
used in the custom RTD. Using the example above, CPRT_COEFB? Would return:

-5.774999E-07

CPRT_COEFC

This command is used for entering a custom RTD into the calibrator. The numeric
value entered after the command will be set as the first coefficient of the polynomial
used by the custom RTD.

For example:

CPRT_COEFC –4.183000E-12 enters –4.183000E-12 as coefficient C.

CPRT_COEFC?

Returns the number that was entered for the third coefficient of the polynomial used
in the custom RTD. Using the example above CPRT_COEFC? Would return:

–4.183000E-12

CPRT_R0

Sets the 0° resistance, R0, in the custom RTD. The value must be entered with a units
label. Refer to the Parameter Units table for assistance.

For example:

CPRT_R0 100 OHM sets R0 to 100 ohms.

CPRT_R0?

Returns the value for the resistance in custom RTD. The above example would return:

1.000000E+02, OHM

DAMP

Turns the dampening function on or off.

For example:

If you send DAMP ON this will turn the dampening function on.