Martel Electronics BetaGauge 321A User Manual

Page 44

Advertising
background image

42

BetaGauge 311A/321A Ranges and Resolutions

Range (PSI)

0.3

1

5

15

30

50

100

300

500

1000

1500

3000

5000

10000

Burst Pressure (PSI)

10

10

100

500

500

1000

1000

2000

2000

10000

10000

10000

10000

15000

Proof Pressure (PSI)

1

3

15

30

60

200

200

600

1000

3000

3000

6000

10000

15000

Engineering Unit

F

actor

P

s

i

1

0.3000

1.0000

5.0000

15.000

30.000

50.000

100.00

300.00

500.00

1000.00

1500.0

3000.0

5000.0

10000.0

bar

0.06894757

0.0207

0.0689

0.3447

1.0342

2.0684

3.4447

6.8947

20.684

34.474

68.947

103.42

206.84

344.74

689.47

mbar

68.94757

20.684

68.948

344.74

1034.2

2068.4

3447.4

6894.8

20684

34474

68948

N/A

N/A

N/A

N/A

kP

a

6.894757

2.0684

6.8948

34.474

103.42

206.84

344.74

689.48

2068.4

3447.4

6894.8

10342

20684

34474

68948

MP

a

.00689476

0.0021

0.0068

0.0344

0.1034

0.2068

0.3447

0.6894

2.0684

3.4474

6.8948

10.342

20.684

34.474

68.948

kg/cm2

0.07030697

0.0211

0.0703

0.3515

1.0546

2.1092

3.5153

7.0307

21.092

35.153

70.307

105.46

210.92

351.53

703.07

cmH2O @ 4°C

70.3089

21.093

70.309

351.54

1054.6

2109.3

3515.4

7030.9

21093

35154

70309

N/A

N/A

N/A

N/A

cmH2O @ 20°C

70.4336

21.130

70.434

352.17

1056.5

2113.0

3521.7

7043.4

21130

35217

70434

N/A

N/A

N/A

N/A

mmH2O @ 4 °C

703.089

210.93

703.09

3515.4

10546

21093

35154

70309

N/A

N/A

N/A

N/A

N/A

N/A

N/A

mmH2O @ 20°C

704.336

211.30

704.34

3521.7

10565

21130

35217

70434

N/A

N/A

N/A

N/A

N/A

N/A

N/A

inH2O @ 4°C

27.68067

8.3042

27.681

138.40

415.21

830.42

1384.0

2768.1

8304.2

13840

27681

41521

83042

N/A

N/A

inH2O @ 20°C

27.72977

8.3189

27.730

138.65

415.95

831.89

1386.5

2773.0

8318.9

13865

27730

41595

83189

N/A

N/A

inH2O @ 60°F

27.70759

8.3123

27.708

138.54

415.61

831.23

1385.4

2770.8

8312.3

13854

27708

41561

83123

N/A

N/A

mmHg @ 0°C

51.71508

15.515

51.715

258.58

775.73

1551.5

2585.8

5171.5

15515

25858

51715

77573

N/A

N/A

N/A

inHg @ 0°C

2.03602

61.081

2.0360

10.180

30.540

61.081

101.80

203.60

610.81

1018.0

2036.0

3054.0

6108.1

10180

20360

N/A - F

or calibrators with firmware version 2.0 and higher

, some units will not be displayed due to limitations on display re

solution. F

or calibrators with earlier versions of firmware, the calibrator

will attempt to display a value, but the display will be excessively noisy to be useful. In all cases, best practice would proh

ibit the use of small engineering units such as mm H2O with high pres-

sure range sensors.

• Proof pressure - maximum allowable pressure without a shift in calibration

• Burst pressure - sensor damaged or destroyed; some risk of personnel injury

• Compound ranges - the data for the 15 PSI range also applies to the -15 to +15 PSI compound range; the data for the 30 PSI ra

nge also applies to the -15 to +30 PSI compound range.

• Absolute ranges - the data for the 15, 30, 50, 100 and 300 PSI ranges also applies to the absolute pressure versions of those

ranges.

Advertising