Martel Electronics BetaGauge 321A User Manual

Page 31

Advertising
background image

29

Indefinite ASCII (IAD)

Any ASCII characters followed by a terminator. For example:

*IDN? returns BETA, 311A/321A, 250, 1.00

12.4.4 Calibrator Status

Error Queue

If an error occurs due to invalid input or buffer overflow, its error code is sent to the error
queue. The error code can be read from the queue with the command FAULT?. The error
queue holds 15 error codes. When it is empty, FAULT? returns 0. The error queue is
cleared when power is reset or when the clear command *CLS is entered.

Input Buffer

Calibrator stores all received data in the input buffer. The buffer holds 250 characters. The
characters are processed on a first in, first out basis.

12.5 Remote Commands and Error Codes

The following tables list all commands, and their descriptions, that are accepted by the
calibrator.

Table 5: Common Commands

Command Description

*CLS

(Clear status.) Clears the error queue.

*IDN?

Identification query. Returns the manufacturer, model number, and firmware
revision level of the Calibrator.

*RST

Resets the calibrator to the power up state.

Table 6: Calibrator Commands

Command Description

CPRT_COEFA

Sets the custom RTD coefficient A

CPRT_COEFA?

Returns the custom RTD coefficient A

CPRT_COEFB

Sets the custom RTD coefficient B

CPRT_COEFB?

Returns the custom RTD coefficient B

CPRT_COEFC

Sets the custom RTD coefficient C

CPRT_COEFC?

Returns the custom RTD coefficient C

CPRT_R0

Sets the custom RTD R0 resistance

CPRT_R0?

Returns the custom RTD R0 resistance

DAMP

Turns Damp on or off.

DAMP?

Returns if DAMP is on/off

DISPLAY

Turns on/off the displays specified in the command

DISPLAY?

Returns which displays are on/off

FAULT?

Returns the most recent error code

Advertising