Martel Electronics PTC8010 User Manual

Page 23

Advertising
background image

23

Command

Description

FAULT?

Returns the error code of an error that has occurred

FUNC?

Returns the current mode of the calibrator

LOCAL

Returns user to manual operation of the calibrator

LOCKOUT

Locks out the keypad of the calibrator, and allows for remote operation only

MEAS

Sets the mode for measuring of the calibrator

OUT

Sets the output of the calibrator

OUT?

Returns the output of the calibrator

REMOTE

Puts the calibrator in remote mode

RTD_TYPE

Sets the RTD type

RTD_TYPE?

Returns the RTD type

RTD_WIRE

Sets the number of wires used by the RTD mode.

RTD_WIRE?

Returns the wire number setting used in the RTD mode

TEMP_UNIT

Sets input/output temperature unit for RTD and TC

TEMP_UNIT?

Returns the temperature unit for RTD and TC

TSENS_TYPE

Sets temperature sensor type.

TSENS_TYPE?

Returns temperature sensor type

VAL?

Returns the measured values

Table 7: Parameter units

Units

Meaning

MV

Voltage in millivolts

V

Voltage in volts

Ohms

Resistance in Ohms

Cel

Temperature in Celsius

Far

Temperature in Fahrenheit

Kel

Temperature in Kelvin

Table 8: Error codes

Error Number

Error Description

100

A non-numeric entry was received where it should be a numeric entry

101

Too many significant digits entered

102

Invalid units or parameter value received

103

Entry is above the upper limit of the allowable range

104

Entry is below the lower limit of the allowable range

105

A required command parameter was missing

107

An invalid CJC_STATE was received

108

An invalid TSENS_TYPE was received