Martel Electronics PTC8010 User Manual

Page 33

Advertising
background image

RTD Type

Range (°C)

Accuracy (°C)

Minimum

Maximum

Measure 4W/Source

Measure 3W

PT385, 200 ohm

-200.0

-80.0

0.6

0.6

-80.0

0.0

0.7

0.7

0.0

100.0

0.7

0.8

100.0

260.0

0.7

0.8

260.0

300.0

0.7

0.8

300.0

400.0

0.9

0.9

400.0

630.0

0.9

1.0

PT385, 500 ohm

-200.0

-80.0

0.2

0.3

-80.0

0.0

0.3

0.3

0.0

100.0

0.3

0.3

100.0

260.0

0.4

0.4

260.0

300.0

0.4

0.4

300.0

400.0

0.4

0.5

400.0

630.0

0.5

0.5

PT385, 1000 ohm

-200.0

-80.0

0.2

0.2

-80.0

0.0

0.2

0.2

0.0

100.0

0.2

0.2

100.0

260.0

0.2

0.2

260.0

300.0

0.3

0.3

300.0

400.0

0.3

0.3

400.0

630.0

0.4

0.4

NI120

-80.0

260.0

0.1

0.2

Cu10

-100.0

260.0

1.3

2.5

Cu50

-180.0

200.0

0.3

0.5

Cu100

-180.0

200.0

0.1

0.3

YSI 400

15.0

50.0

0.1

0.1

33