Martel Electronics PTC8010 User Manual

Page 24

Advertising
background image

24

Error Number

Error Description

110

An unknown command was received

111

An invalid RTD or TC parameter value was received

112

The serial input buffer overflowed

113

Too many entries in the command line

114

The serial output buffer overflowed

115

Output is overloaded

5.5 Entering Commands

Commands for the calibrator may be entered in upper or lower case. There is at least one
space required between the command and parameter, all other spaces are optional. Almost
all commands for the calibrator are sequential, any overlapped commands will be indicated
as such. This section will briefly explain each of the commands and describe their general
use, which will include any parameters that may be entered with the command as well as
what the output of the command is.

5.5-1 Common Commands

*CLS

Clears the ESR, the error queue and the RQS bit. Also terminates all pending operations.
When writing programs, use before each procedure to avoid buffer overflow.

*ESE

Loads a byte into the Event Status Enable register. The command is entered with a decimal
number that, when converted to binary, enables the right bits in the Event Status Register. For
example:

*ESE 133

When 133 is converted to binary it is 10000101. Bits 7, 2, and 0 will be enabled.

*ESE?

Returns the contents of the Event Status Enable register. The value returned is a decimal. For
example, if the register has the following settings:

10000101

than the value returned will be 133.

*ESR?

Returns the contents of the Event Status Register in decimal form. For example:

If the ESR contains 10111001, *ESR? will return 185.