*IDN?

Returns the manufacturer, model number, and firmware revision of the Calibrator. For
example:

*IDN? will return MARTEL, PTC8010, 250, 1.00

*OPC

Enables the Operation Complete setting in the ESR. This setting makes it possible to check if
an operations is complete after it has been initialized.

For example this operation could be used with the command TRIG.

*OPC?

Returns 1 when all operations are complete, and causes program execution to pause until all
the operations are complete. For example:

TRIG ; *OPC? will return a 1 when the pulse train initiated by TRIG is complete.

*RST

Resets the state of calibrator to the power-up state. All subsequent commands are held off
until the execution of the command is complete.

*SRE

Loads a byte into the Service Request Enable register. A decimal number must be entered,
which when converted to binary, corresponds to the correct settings.

For example:

*SRE 8 enters the binary number 00001000 to the SRE. This enables bit 3. Bit 6 is not
used.

*SRE?

Returns a byte from the SRE. The byte is returned in decimal format. For example:

If 40 is returned, bits 5 and 3 are enabled.

*STB

Returns the status byte in decimal form from the Serial Poll Status Byte. For example;

If 72 is returned, bits 6 and 3 are enabled.

*WAI

Prevents further remote commands from being executed until all previous commands are
executed. For example:

OUT 100°C ; *WAI ; OUT 200°C will output 100°C and wait until output settles, then the
second output command will be processed.

25