6. Specifications

One year at 23°C ± 5°C, unless specified otherwise. Outside of this range the stability of the
measurements is ± 0.003% of FS/°C.

Table 9: General Specifications

Operating Temperature

-10°C to 50°C

Storage Temperature

-20°C to 60°C

Power

4 X AA batteries; Alkaline or optional rechargeable

Battery Life

Approximately 20 hours

Serial Communications

ASCII; RS232 or USB depending on the communications cable used

CE - EMC

EN61326: 2006, ANNEX B Criterion A

Safety

CSA C22.2 No. 1010.1: 1992

Size (with holster)

8.5” x 4” x 2” (21.6cm x 10.2cm x 5.1cm)

Weight

Approximately 28oz. (0.8kg)

Table 10: Resistance Measurement

Range

Accuracy (% of reading ± floor)

Ohms low

0.00

Ω - 400.0Ω

0.025% ± 0.05

Ω

Ohms high

401.0

Ω - 4000.0Ω

0.025% ± 0.5

Ω

Table 11: Resistance Source

Range

Excitation Current

Accuracy (% of reading ± floor)

Ohms low

5.0

Ω - 400.0Ω

0.1mA - 0.5mA

0.025% ± 0.1

Ω

5.0

Ω - 400.0Ω

0.5mA - 3mA

0.025% ± 0.05

Ω

Ohms high

400

Ω - 1500Ω

0.05mA - 0.8mA

0.025% ± 0.5

Ω

1500

Ω - 4000Ω

0.05mA - 0.4mA

0.025% ± 0.5

Ω

Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <= 5ms.

Table 12: Thermocouple Measurement/Source

Range

Accuracy (% of reading ± floor)

Read (mV)

-10.000mV - 75.000mV

0.02% ± 10

μV

Source (mV)

-10.000mV - 75.000mV

0.02% ± 10

μV

Maximum current output in voltage ranges is 1mA with an output impedance of <= 1

Ω