Specifications – Martel Electronics PTC8010 User Manual
Page 30

30
6. Specifications
One year at 23°C ± 5°C, unless specified otherwise. Outside of this range the stability of the
measurements is ± 0.003% of FS/°C.
Table 9: General Specifications
Operating Temperature
-10°C to 50°C
Storage Temperature
-20°C to 60°C
Power
4 X AA batteries; Alkaline or optional rechargeable
Battery Life
Approximately 20 hours
Serial Communications
ASCII; RS232 or USB depending on the communications cable used
CE - EMC
EN61326: 2006, ANNEX B Criterion A
Safety
CSA C22.2 No. 1010.1: 1992
Size (with holster)
8.5” x 4” x 2” (21.6cm x 10.2cm x 5.1cm)
Weight
Approximately 28oz. (0.8kg)
Table 10: Resistance Measurement
Range
Accuracy (% of reading ± floor)
Ohms low
0.00
Ω - 400.0Ω
0.025% ± 0.05
Ω
Ohms high
401.0
Ω - 4000.0Ω
0.025% ± 0.5
Ω
Table 11: Resistance Source
Range
Excitation Current
Accuracy (% of reading ± floor)
Ohms low
5.0
Ω - 400.0Ω
0.1mA - 0.5mA
0.025% ± 0.1
Ω
5.0
Ω - 400.0Ω
0.5mA - 3mA
0.025% ± 0.05
Ω
Ohms high
400
Ω - 1500Ω
0.05mA - 0.8mA
0.025% ± 0.5
Ω
1500
Ω - 4000Ω
0.05mA - 0.4mA
0.025% ± 0.5
Ω
Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <= 5ms.
Table 12: Thermocouple Measurement/Source
Range
Accuracy (% of reading ± floor)
Read (mV)
-10.000mV - 75.000mV
0.02% ± 10
μV
Source (mV)
-10.000mV - 75.000mV
0.02% ± 10
μV
Maximum current output in voltage ranges is 1mA with an output impedance of <= 1
Ω