Agilent Technologies E5810A User Manual

Page 126

Advertising
background image

126

Chapter 4

Troubleshooting Information
Checking PC Client Connections

E5810 appears
to be hung

There are several possible causes. For all of the following, use of client
timeouts and/or E5810 timeouts (the I/O timeout and LAN timeout
configuration values) can be used to ensure that the E5810 will not
“hang” indefinitely.

„

If you determine that the E5810 is truly hung, you may be able to

free it by terminating a particular client connection at the client,
thus freeing the resources to allow the E5810 to proceed.
Otherwise, you can press the

Clear ALL Pending Operations

button on the

Find and Control Instruments Connected to your

E5810

Web page or you can reboot the E5810.

„

The network connection to the E5810 has been broken.

„

The E5810 may be attempting a transaction to a non-existent or

powered-off device. Check the address used, as well as the
status of the device.

„

The E5810 is busy performing a very long transaction to a (slow)

device, or it is waiting for input from a device.

„

A client may be waiting for access to a device/interface locked by

another client. Check for possible deadlock situations and, if
possible, make sure the clients owning locks and the connections
to those clients are still up and functioning. You can check the
current session by clicking the

Session Status

icon to display the

E5810 LAN/GPIB Gateway Status

Web page.

Message

Description/Actions