14

5.3

Calibration with Pressure

5.3.1

Zero Point Calibration

5.3.2

Span Calibration

Calibration of the measurement range (span).

Determine if the pressure to be used as the zero point (P 0 %), is present
at the transmitterer diaphragm before calibration.

Ensure that the pressure to be used as the span end-point (P 100 %) is
present at the transmitter diaphragm.

A change in the zero point has no effect on the calibrated span.
However, if the span end-point is higher than the peak value of the sen-
sor’s nominal pressure range, then span end-point is fixed at this peak val-
ue and the span is reduced accordingly.
A change in the span setting has no effect on the zero point. The zero point
and span end-point must fall within the sensor’s nominal pressure range.

meas

Zero point calibration

Calibration Value

meas

Calibration of span

Calibration Value