NOSHOK 756 Series Digital Pressure Transmitters User Manual

Page 19

Advertising
background image

16

Example:

A pressure transducer with 0 ... 400 mbar (nominal pressure) is to be programmed.
The liquid’s surface (with a density of 1) is 1 m above the diaphragm at the zero point.
The maximum (span end-point) should be 3 m.

This means that the output must be set to 12 mA during programming.

A change in the zero point has no effect on the adjusted span.
However, if the span end-point is higher than the peak value of the
sensor’s nominal pressure range, then the span end-point is fixed at
this peak value and the span is reduced accordingly.
A change in the span setting has no effect on the zero point. The zero
point and span end-point must fall within the sensor’s nominal pres-
sure range.

A test / correction of the zero point is suggested after adjusting the
span in order to maintain optimum accuracy.

I

c al c

pressure difference (span) (300 mbar -100 mbar)

Sensors nominal pressure 400 mbar

---------------------------------------------------------------------------------------------------------------------------------- 16 mA 4 mA

+

12 mA

=

=

Enter
Value

Access
Function

... to Imeas = Icalc

Activate

Keypad

(2 s)

Calculate

Calibration Value

Calibration of span, without display, without pressure

current value decreases

Confirm
Calibration Value

... to Imeas = Icalc

current value increases

Icalc =

Important

Advertising